Portfolio · Diagnostic tests

Sepsis/Meningitis

EuSep Screen® Lattanti

REF

9144

Description

Real time PCR test for the detection of bacterial DNA belonging to Klebsiella pneumoniae, Escherichia coli, Streptococcus agalactiae (Streptococcus B or GBS) and Listeria monocytogenes.

Packaging

8 test

Certification

“Sepsis is a rare complication of an infection, the consequences of which can be very serious and potentially fatal. It consists of an excessive inflammatory response of the body to a generalized infection that damages tissues and organs, compromising their functioning” – Istituto Superiore di Sanità (ISS) – 2019
Molecular diagnostic tools enable the pathogens identification in a short time and therefore they can limit the use of broad-spectrum antibiotics, which may be un-effective. EuSep Screen® Lattanti is able to detect non-viable pathogen as a consequence of antibiotics administration before the sample’s collection

Summary

Technology

Real Time PCR technology

Target

  • Group B Streptococcus
  • Klebsiella pneumoniae
  • Escherichia coli
  • Listeria monocytogenes

Genetic region

  • gene phoE (Klebsiella pneumoniae all serotypes)
  • gene uidB (Escherichia coli, all serotypes)
  • gene sip (Streptococcus agalactiae, all serotypes)
  • gene iap (Listeria monocytogenes, meningites related strain)

Mastermix number

3

Internal control

Exogenous

Fluorophores

  • FAM
  • JOE

Limit of Detection (LoD)

2 copies/µl (10 copies/reaction)

Analytical specificity

A wide range of pathogens was used to test the assay. No cross-reactivity was found.

Diagnostic sensitivity

100 %

Diagnostic specificity

100 %

Specimen

  • Whole blood
  • CSF
  • Pleural fluid
  • Synovial fluid
  • Other normally sterile human biological material from newborn

Validated thermalcyclers

  • Applied Biosystems™ 7300 and 7500, 7500 Fast e 7500 Fast Dx
  • Step One e Step One Plus Real-time PCR Instrument system (Thermo Fisher Scientific)
  • CFX96 Touch, CFX Connect, CFX Opus 96 Real-time PCR detection system (Bio-Rad)
  • Eppendorf RealPlex
  • Qiagen Rotor-Gene 3000/6000